Three things change sound speed in CMP slurry. Only one of them is H2O2.
Our ICPT 2026 paper shows how the Rhosonics Chemical Concentration Meter separates them: inline, up to ten times a second, within ≈ ±0.03 wt% of titration from 20 to 28 °C.
Feasibility study presented at ICPT 2026, from the paper Inline Real-Time Monitoring of Hydrogen Peroxide Concentration in CMP Slurries Using Ultrasonic Technology for Advanced Process Control by Raymond Maas and Stefan Seuleanu, Rhosonics.
H2O2 changes the speed of sound. That makes it measurable inline.
Titration gives you one sample, measured in a lab, at one temperature. Sound speed can be read in the slurry loop, continuously. The catch: H2O2 isn’t the only thing that moves it.
| Titration | Rhosonics CCM | |
|---|---|---|
| Where it measures | A sample, taken to the lab | Inline, in the recirculation loop |
| How often | When someone takes a sample | Up to 10 times a second |
| Temperature | The lab’s, not the loop’s | Measured and compensated live |
| When you know | After sampling and analysis | Immediately, in your control system |
Titration still matters. It provides the reference samples that calibrate the CCM, and it checks it.
Temperature moves sound speed too, and not by the same amount at every concentration.
Sound speed shifts strongly with temperature, and the size of that shift depends on how much H2O2 is in the slurry. A single temperature correction, tuned for one concentration, is wrong for all the others.
Its model lets the temperature correction change with the H2O2 level, so the reading holds as your loop warms and cools.
As slurry ages, sound speed drifts even when H2O2 doesn’t.
We circulated a copper CMP slurry at about 1 wt% H2O2 for more than two months to age it. Its base properties changed, and a sound-speed reading drifted with them. Possible causes:
- Water lost to evaporation
- H2O2 decomposing
- Changes in slurry stability
Conductivity follows the base slurry, not the H2O2. Adding it to the model separates the two. Uncompensated drift: ≈ +0.028 wt% in 48 hours. Compensated: on target.
With both traps handled, the reading stays within ≈ ±0.03 wt% of titration.
One copper CMP slurry, three titrated H2O2 levels, loop temperature swept from 20 to 28 °C.
- 0.9618 wt%, 20 to 28 °C≈ ±0.03
- 1.9780 wt%, 20 to 28 °C≈ ±0.03
- 2.8552 wt%, 20 to 26 °C≈ ±0.02
- Spread at steady temperature≈ ±0.015
- Low concentrations told apart0.04 · 0.07
ICPT 2026 poster session. Wednesday 21 October, 13:10 to 15:10.
The method works. Next, we calibrate it to your slurry.
CMP slurries differ in abrasives, additives and oxidiser levels, so the CCM’s model is modular. Its coefficients are fitted to your formulation from reference samples at several H2O2 levels and temperatures.
Four measurements, taken at one point in the line
The H2O2 signal.
From the echo at the sensor face, with sound speed.
Follows the base slurry, to separate ageing from H2O2.
At the same point, with a correction that changes with concentration.
- 01 · Install
Inline, at the mixing tank
In the recirculation loop, before the slurry reaches the polishers.
- 02 · Calibrate
To your slurry
With titrated reference samples at several H2O2 levels and temperatures.
- 03 · Monitor
Continuously
On the display and in your control system, compensated for temperature and slurry ageing.
CMP slurry is new for us. Fab chemistry isn’t.
Rhosonics concentration meters have measured developers, etchants and process chemicals inline in display and chip plants since 2009. The CMP work builds on that.
Chemistries we already measure inline
Display makers and chemical suppliers in Asia
Display makers in Korea and China
Chemical producers in Taiwan
Semiconductor packaging plants in Taiwan
Chemical and nuclear fuel producers in Europe
Chemical producers in Asia and Europe
DS Linetech works with us to solve difficult chemical concentration measurements for some of the biggest flat-panel display and semiconductor companies in South Korea.
Track record
- 2009Analysers in flat-panel display plants
- 2023Partnership with S3 Alliance, German maker of automated wet-processing tools
- 2024First CMP slurry paper, ICPT Wiesbaden Read it
- 2025CMP slurry monitoring presented at SEMICON China
- 2026Calibrated H2O2 results at ICPT Seoul
Let’s talk about your slurry, at ICPT or after.
At the show? Find Stefan at our poster on Wednesday 21 October, or leave your details and we’ll arrange a meeting in Seoul. Can’t make it? Tell us what you’re running, CMP slurry or any other fab chemical, and we’ll set up a call.
Poster session, SeoulPresented by Stefan Seuleanu
Wed 21 Oct, 13:10 to 15:10
Stefan SeuleanuHead of R&D and Product, presenting
s.seuleanu@rhosonics.com
Раймонд МаасChief Technology Officer, lead author
r.maas@rhosonics.com
DS LinetechOur partner in Korea, solving difficult concentration measurements with us for the country’s leading display and chip makers.
Talk to a specialist
At ICPT or afterwards. A Rhosonics application specialist will reply.